SECTION 5—WHEELCHAIR OPERATION

1.Hold the light GREY collar portion of the joystick connector with one hand and the controller connector on the wheelchair in the other and align them.

2.Lightly push to engage the joystick connector and the controller connector.

Light Grey Collar

Controller

Connector

Joystick Connector

FIGURE 5.11 Disconnecting/Connecting

the Joysticks - SPJ+ Joysticks

MPJ+ Joysticks

NOTE: For this procedure, refer to FIGURE 5.12.

Disconnecting

1.Pull the latch away from the joystick connector.

2.Disconnect the joystick connector from the remaining connectors.

Connecting

NOTE: Ensure the gaskets are installed in the top connector cap and between network connectors.

1.Ensure the latch is pulled away from the network connector.

2.Connect the network connector to the other connectors.

3.Top and Bottom Connectors ‐ Install connector caps onto the network connector.

4.Push the latch in to secure the network connectors and caps.

5.If necessary secure excess cable using tie‐wraps.

Top Connector

Cap

Latch

Gasket

Network

 

Connectors

 

Controller

Gasket

Cable

 

Bottom

Gasket

Connector Cap

FIGURE 5.12 Disconnecting/Connecting

the Joysticks - MPJ+ Joysticks

Part No 1154294

45

TDX® SI, TDX SI-HD

Page 45