Troubleshooting
2-14 March 2001
<27> BAD_PATCH If the Pattern Recognition comparison window is set to 0 or 1, this error
may result when a single patch on the strip is bad due to dirt, smears,
creases, etc. (See explanation of CW command for more information. )
<30> BAD_STRING_ LENGTH A strip definition string received as part of the DS (Define Strip) RCI
command did not contain the correct number of characters.
<31> BAD_CHARACTER A non-printable character was received in one of the name fields (strip
name or pass name). Printable characters are in the range 0x20 to 0x7E,
inclusive.
<32> BAD_MEAS_TYPE The value for the measure type field was greater than 3. Valid values are
0 through 3.
<33> BAD_COLOR The value for the color field was greater than 9.
<34> BAD_STEPS The value for the number of steps was greater than 72.
<35> BAD_STOP_LOC ATION The value for the guide setti ng was less than 4 or greater than 30.
<36> BAD_OUTPUT_TYPE The value for the output type was greater than 5.
<37> MEMORY_ERROR The definition could not be stored.
<38> BAD_N_FACTOR The n-factor (used in Dot calculations) is either less than 0500 (0.500) or
greater than 4500 (4.500).
<39> STRIP_DOESNT_EXIST The SD command was issued and the requested strip does not exist.
<3A> BAD_MIN_MAX_VALUE The value for the min/max field was greater than 4.
<40> BAD_SERIAL_NUMBER An attempt was made to redefine one of the stored serial numbers with a
number containing one or more invalid digits.
Code Message Explanation