Installing and Using the Centralized Densitometer Applica tion
4-12 March 2001
Measuring Densities During a Calibration CycleIMPORTANT: To use the centralized densitometer appl ication for measuring
densities for a calibration cycle, the density source of “ce ntralized
densitometer” must have been selected in the edit config uration
dialog for the device being calibrated; see “Editing Informati on on
the Density Source Tab” on page 3-28.
1. Open the Centralized Densitometer Application. The display shows the
window for the Centralized Densitometer Application.
2. Highlight the printer.
3. Click the Measure icon (or select Operations->Measure) to measure the
target images.
4. At the densitometer, measure the densities (see “Using the Densitometer to
Read Densities” on page 1-7).
When the densities are measured, the status bar on the Centrali zed
Densitometer Application window displays “Ready.”
The density measurements are automatically sent to the LED printe r’s host
computer.