Calibrating the Printer
1-6 March 2001
Obtaining DensitiesThere are three methods for obtaining densities:
•from a file
•from a densitometer
•from a centralized densitometer (NT only and if the Centralized Densitometer
application is installed on your network)
The contents of the calibration configuration file determine wh ich method you use
(see “Editing the Device Configuration” on page 3-25).
Obtaining Densities from a File
If the configuration settings include obtaining the density d ata from a file, the
display shows the Waiting for Density File dialog box.
If the configuration settings do not include obtaining the de nsity data from a file,
go to either:
•“Using the Densitometer to Read Densities” on page 1-7
•“Obtaining Densities from the Centralized Densitometer (NT Only)” on
page 1-8
NOTE: To create a density file, see “Creating a Density File for Use with Calibration”
on page 3-36.
1. Click OK to continue with the calibration process.
2. Go to “Completing the Calibration” on page 1-10.