Calibrating the Printer
March 2001 1-5
2. Click Yes.
The display shows the Choose Start LUT dialog box.
NOTE: The highligh ted LUT in the dialog box above is the most recent
calibration LUT.
3. Click either Load Selected LUT or Load Custom LUT (or Cancel to cancel
the calibration process).
If you click Load Selected LUT, the system automatically downloads the
highlighted LUT to the printer to create a test print. When the test pr int has
been sent, the Scanning Processed Print with Densitometer status icon on
the LED Printer Calibration window highlights.
Continue with Obtaining Densities on page 1-6.
4. If you selected Load Custom LUT in step 3:
From the Open dialog box, select or type the name of the LUT file you want
and click Open.
The system automatically downloads the LUT file to the printe r.
When the test print has been sent to the printer, the Scanning Process ed
Print with Densitometer status icon on the KODAK LED Printer Calibra tion
window highlights.