Leica 106Z user manual Stereo-fluorescence module Focusing drive base of stand

Models: 106Z

1 24
Download 24 pages 42.62 Kb
Page 21
Image 21

Stereo-fluorescence module

Focusing drive base of stand

Connect manually-operated or motor-driven focusing drive, complete with column, to base, as described in user manual M2-105-0EN..

Before using the motor focus, you must read the safety directions in the accompanying user manual M2-267-104.

For the Leica stereomicroscope models MS5, MZ6, MZ8, MZ12, MZAPO with manual focusing drive, fit the micro­scope in

the upper position on the focusing drive ­(refer to user manual for stereomicroscope)..

Fluorescence module stereomicroscope

Assemble in the following­ order without double-iris dia- phragm: Optics carrier, fluorescence module, bin­ocular tube or phototube

Assemble in the following­ order with double-iris diaphragm:

Optics carrier, double-iris diaphragm, fluorescence module, binocu­lar­ tube or phototube­

Align the fluorescence module on the optics carrier so that

the adapter­ for the lamp housing points backwards at about 45°

Continue assembling in accordance with the user manual for the stereomicroscope (see section “Fitting accessory tubes”)

Video-/phototubes

The fluorescence illumination­ is directed along the left-hand beam path and could cause hot-spots in the image.. Therefore:

Position the HU phototube­ over the right­ -hand beam path.

Changing the filter set

Removable filter sets that can hold various filters are available for the fluorescence module.. A filter set consists of a filter carrier with barrier filter and of a filterholder with excitation filter..

If you wish to change the filter set while you are in the middle of a task, leave the illuminator switched on so that you will

not have to wait for the lamp to cool (page 12) before continuing..

3

Important: With the illuminator on, insert the filterhol- der with diaphragm into slide-in unit 3 to avoid being suddenly exposed to excessively bright light (page 22).

Remove the binocular tube..

Loosen the 3 socket-head screws..

Remove the cover..

Loosen the fastening screw and take out the filter carrier..

Insert the desired filter carrier.. Fit both of the two guide pins in the fluorescence module into the holes on the filter carrier..

Tighten the fastening screw securely..

Replace the cover and tighten the 3 socket-head screws securely..

Replace the binocular tube..

Insert a suitable exciter filter (page 22)..

Take out the filterholder with diaphragm..

Re-adjust the light source (page 15)..

21

Page 21
Image 21
Leica 106Z user manual Stereo-fluorescence module Focusing drive base of stand