Servicing and repairs

Repairs may only be carried out by Leica-trained service technicians or by technical personnel authorized to do so by the person in charge of the instrument

Only original Leica spare parts may be used

Unplug the power cable before opening the supply unit.. Touching live circuits can cause injury..

Legal requirements

Adhere to general and local regulations relating to accident prevention and environmental protection..

Conformity with European Community directive

Leica stereo-fluorescence systems and their accessories are constructed in accordance with the latest technologies and are provided with a statement of conformity with EC requirements..

Supply unit

Pull the power plug of the supply unit out from the power soc- ket:

when assembling and dismantling the lamp housing

before opening the lamp housing

when changing the mercury-vapour burner and other components such as the heat-­protection filter or the collector

when servicing the supply unit

Lamp housing

Never open the lamp housing while the burner­ is switched on (risk of explosion, UV irradiation, dazzling)

Always allow the lamp housing to cool for at least 15 minutes before opening it (risk of burner exploding)

Never cover the ventilator slots on the lamp housing (risk of fire)

 

Mercury-vapour lamp

 

 

• Read the user manual and safety directions provided by the

 

manufacturer of the mercury-vapour burner, and particularly

 

­those relating to its breakage and to the ­release of mercury

Light source: Safety regulations

• Before transporting the equipment, remove the mercury-vapour

 

burner and place it in its original packaging.. Use the transport

Safety measures introduced by manufacturer

peg in place of the burner, to secure movable parts within the

lamp housing

 

– UV protection screen in front of the specimen plane prevents

• When the mercury-vapour burner has reached the end of its

direct UV radiation from reaching the eyes..

nominal working life as indicated by the manufacturer’s

– Dummy filter carriers in the unoccupied positions of the rapid

information and the time counter on the supply unit, change the

discoloured burner (increased risk of explosion)­

filter changer prevent direct UV radiation from reaching

• Leica declines all responsibility for injury and damage

the eyes

– UV filters in the observation beam paths protect the eyes

resulting from exploding, incorrectly­

-fitted or incorrectly-used

against UV radiation..

mercury-­vapour burners..

 

Stray-light protection on the underside of the lamp housing

 

protects the user's hands against UV radiation..

 

Warning

UV can damage your eyes.. Therefore:

Never look at the light spot on the specimen plane without the UV protection screen in position

Never look into the eyepieces unless there is an excitation filter in the beam path

Always keep dummy filter carriers in the unoccupied positions of the rapid filter changer (for MZ10 F)

Do not place the specimen on a white or highly-reflective background..