3.Instrument components and specifications

3.3 Technical data General

Approvals:

The instrument-specific marks are located next to

 

the identification label.

Nominal supply voltages:

100 / 120 / 230 / 240 V AC ±10%

Nominal frequency:

50/60 Hz

Power draw:

70 VA

Protective class :

I

Power fuses:

2 x T 1.0 A UL listed

Pollution degree: :

2

Overvoltage installation :

II

Maximum heat emission:

70 J/s

Operating temperature range:

+10°C to +35°C

Temperature range during storage:

+5°C to +55°C

Relative humidity:

max. 80% non-condensing

Humidity during storage:

< 80%

according to IEC-1010, UL 3101, EN 61010

Dimensions and weight

Basic instrument

Width (including handwheel):

413 mm

Width (excluding handwheel):

300 mm

Depth (including waste tray):

618 mm

Height (total):

305 mm (with tray on the hood)

Working height (knife blade):

100 mm (measured from the base plate)

Working height (knife blade):

168 mm (measured from the table)

Weight (without accessories):

approx. 37 kg

Control panel

 

Width:

94 mm

Depth:

164 mm

Height:

50 mm

Height (in inclined position):

81 mm

Weight (net):

approx. 0.450 kg

14

Instruction manual V 1.4 – 02/2007

Page 14
Image 14
Leica RM2245 instruction manual Technical data General, Dimensions and weight Basic instrument, Control panel