5. Operation
Configuring the retraction settings
•To call up the retraction settings, press the MENU MODE and CLEAR
+buttons simultaneously.
•The current set value is displayed as a
•Select the desired retraction value.
or | The retraction value can be adjusted in increments of 5 μm to a maxi- |
| mum of 100 μm using the - buttons on the control panel. |
•To exit the retraction settings, press MENU MODE.
A retraction movement takes place by the newly selected value after each section.
4 | • While the specimen is in retraction, the yellow LED on the RETRACT |
| |
| display (4) lights up. |
Switching off the specimen retraction
•To call up the retraction settings, press MENU MODE and CLEAR simultaneously.
+
•To switch off retraction, press the button on the control panel until the display indicates "OFF".
•To exit the retraction settings, press MENU MODE.
When retraction is switched off, the specimen is not retracted. The yellow LED (4) of the RETRACT indicator does not light up.
4
26 | Instruction manual V 1.4 – 02/2007 |