6.Optional accessories

6.9 Cold light source

1

7

Fig. 58

6.10 Fiber-optic light guide

6

Fig. 59

5

8

2

4

3

Fig. 60

The cold light source may be operated at the power supply voltage specified on the identification label (on the underside of the instrument) only.

Connect the power cable (1) to the connect- ing jack on the rear of the device.

Plug the power plug into the power socket.

Carefully read the instruction manual provided separately before use!

Attach the fiber-optic light guide (2) to the adapter of the magnifying lens and secure it in place in the bore (6) using the screw (5).

Pull off the protective cap (3) from the end of the fiber-optic light guide and keep it in a safe place.

Insert this end into the opening provided (4) until it engages with a slight click.

Switch on the Cold light source using the power switch (7).

Remove the protective caps (8) and align the light guide with the specimen.

Fig. 62, Leica rotary microtome with installed optional accessories: magnifying lens and fiber-optic light guide

56