8.2.3 Lubricating the instrument

Once a month, lubricate the following parts with the included drive parts oil No. 405 (1 - 2 drops are enough).

Instrument and specimen holder:

The clamping key (95) of the quick clamping system.

The T-piece (55) on the microtome base plate.

Guide rails (96) for the knife holder base on the microtome base plate.

The T-piece on the knife holder base (56).

Knife holder E:

Clamping lever (11) for the lateral displace- ment.

T-piece (88) and guide (97) of the knife holder for the lateral displacement.

Clamping lever (10) for the blade.

Knife holders N and NZ:

Sliding surfaces of the finger guard (8).

Knurled nuts (73) for measuring height adjustment.

Universal cassette clamp:

Shaft (90) and flat coil spring (91) of the clamping lever of the cassette clamp. After lubrication, move the lever back and forth several times.