6. Optional accessories

6.2.6 Super mega-cassette clamp

2

4

5

1

3

Fig. 35

Assembly of the super mega-cassette clamp

The super mega-cassette clamp should preferably be used with the rigid fixture for specimen clamps.

To do so, proceed as follows:

Remove the rubber ring (5) only after having attached the cassette clamp on the object head!

Screw the rigid fixture for specimen clamps (4) onto the object head (3): Remove the screw (1), place the fixture for specimen clamps (4) onto the object head (3) from the front and tighten the screws (2) with an Allen key size 3.

Next, insert the screw (1) from the side and briefly tighten it with an Allen key size 4.

Insert the super mega-cassette clamp from the side on the left into the dovetail guide of the rigid fixture for specimen clamps and tighten screw (1).

If the directional fixture for specimen clamps is used with the rigid knife holder base, the orientation must be in position "0" and the cover for backlighting illumination must be detached. (Danger of collision if not observed!)

NEVER use the super mega-cassette clamp with backlighting illumination!

Leica RM2245

43

Page 43
Image 43
Leica RM2245 instruction manual Super mega-cassette clamp