4.Startup
Checking the voltage (continued)
22
21
25
23 24 21
27 26
The voltage selector is located above the main power switch, on the
•Insert a small screwdriver into the cutout (25) and carefully pry out the insert.
•Remove the voltage selector housing (21) together with the fuses (23). Remove the voltage selector block (24) (white) and insert it again such that the correct local voltage is displayed in the viewing window (22).
•Reinsert the voltage selector housing with the block and fuses and push it in until it engages (audible click).
Fig. 10
14 | 17 | 15 |
16
Fig. 11
Connecting the control panel
The connecting cable (15) of the control panel is fixed to the microtome. It may not be discon- nected.
•Insert the plug (14) of the connecting cable (15) into the connecting jack (16) on the rear of the control panel.
•To fasten the plug, tighten the two screws (17).
Leica RM2245 | 19 |