APPENDIX B

SERVICE LOG ERROR MESSAGES

Table B-1.Operational Error Messages (1 of 2)

Error Message

Description

Corrective Action

 

 

 

ATTENUATOR UNAVAILABLE

Option 6 Port 2 Test Step Attenuator

Install Option 6 Step Attenuator,

 

is not installed.

 

 

 

 

DIFFERENT H/W SETUP. RECALL

Model and/or options is (are) different

Reconfigure system to duplicate the

ABORTED

from the recalled setup.

hardware setup that was used to store

 

 

the saved data.

 

 

 

DIFFERENT S/W VERSION, RECALL

Saved state not compatible with soft-

Load compatible software (S/W) ver-

ABORTED

ware version or options.

sion and retry.

 

 

 

FREQUENCIES HAVE REACHED

Frequencies being defined in Multiple

Redefine frequencies to not exceed

UPPER LIMIT

Source mode have reached upper lim-

limits of Sources.

 

its of Sources.

 

 

 

 

MEMORY LOCATION CORRUPTED

Requested memory location is cor-

None. If problem reoccurs after stor-

 

rupted.

ing a new setup, contact ANRITSU

 

 

Customer Service.

 

 

 

NO BANDS ARE STORED

No frequency bands have been de-

Define and store frequency bands to

 

fined and stored.

turn on Multiple Source mode.

 

 

 

NO STORED MEMORY DATA

No data is stored in memory for dis-

Store or re-save measurement data.

 

play or trace math.

 

 

 

 

OPTION NOT INSTALLED

Selected an option that is not in-

None.

 

stalled.

 

 

 

 

OUT OF CAL RANGE

Entered values out of the selected

Change calibration range or re-enter

 

calibration range.

values that are within the current

 

 

range.

 

 

 

OUT OF H/W RANGE

Entered value is out of the instru-

Re-enter values that are within range.

 

ment’s hardware range.

 

 

 

 

OUT OF RANGE

Entered value is out of range.

Re-enter values that are within range.

 

 

 

RECEIVER OUT OF RANGE BY

Equation defined in Multiple Source

Redefine frequency.

EQUATION

mode places receiver frequency out of

 

 

range when attempting to store band.

 

 

 

 

SOURCE 1 OUT OF RANGE BY

Equation defined in Multiple Source

Redefine frequency.

EQUATION

mode places Source 1 frequency out

 

 

of range when attempting to store

 

 

band.

 

 

 

 

SOURCE 2 OUT OF RANGE BY

Equation defined in Multiple Source

Redefine frequency.

EQUATION

mode places Source 2 frequency out

 

 

of range when attempting to store

 

 

band.

 

 

 

 

373XXA MM

B-5

Page 161
Image 161
Anritsu 373XXA manual Service LOG Error Messages, Error Message Description Corrective Action

373XXA specifications

The Anritsu 373XXA series is a notable line of microwave and RF signal analyzers designed for comprehensive analysis and testing of various wireless communication technologies. These analyzers are specifically crafted to meet the demands of modern communication systems, making them ideal for engineers and technicians working in wireless research, development, and manufacturing.

One of the main features of the Anritsu 373XXA series is its wide frequency range, which typically spans from as low as 5 kHz to several GHz, accommodating a vast array of signal types and applications. This extensive frequency coverage allows users to test and validate all manner of systems, from LTE and 5G networks to satellite communications, yielding accurate and reliable measurement results.

The device employs advanced measurement technologies that provide in-depth analysis capabilities. Among these technologies is the ability to perform both time and frequency domain measurements, allowing users to visualize and diagnose signal integrity issues effectively. In addition, the 373XXA series incorporates vector signal analysis, which is crucial in evaluating complex modulated signals, commonly found in today’s sophisticated wireless communication protocols.

Another distinguishing characteristic of the Anritsu 373XXA is its user-friendly interface. It features a large, high-resolution display that enables easy navigation through the analyzer's functions and measurements. The intuitive design is further enhanced by customizable settings and measurement configurations that streamline the workflow, making it accessible for both novice and experienced users.

The 373XXA series also supports various digital and analog signal demodulation techniques, ensuring compatibility with an array of communication standards. Furthermore, users can benefit from its built-in signal generation capabilities, which allows for end-to-end testing of devices and systems under realistic operational conditions.

Portability is another key feature of the Anritsu 373XXA series. The compact design and lightweight structure facilitate easy transport, which is particularly advantageous for field applications where mobility is critical.

Overall, the Anritsu 373XXA series stands out as a versatile and powerful tool for professionals in the RF and microwave field, thanks to its extensive frequency range, advanced measurement technologies, user-friendly interface, and portability. Whether developing new wireless standards or troubleshooting existing systems, the 373XXA delivers precision and reliability essential for today’s dynamic communication landscape.