Anritsu 373XXA manual A9 PCB Sram Battery Remove and Replace

Models: 373XXA

1 202
Download 202 pages 15.37 Kb
Page 124
Image 124

A9 PCB SRAM BATTERY

REMOVE AND REPLACE

8-6A9 PCB SRAM BATTERY This paragraph provides instructions for removing and replacing the SRAM Battery located on the A9 Main Processor PCB assembly.

Preliminary:

￿Switch the 373XXA power off. Remove the power cord.

CAUTION

The SRAM battery incorporates flammable materials such as lithium and organic solvents. If lithium batteries, such as this one, are mistreated or incorrectly handled, they may burst open and ignite. This can result in possible in- jury or fire. When handling the SRAM battery, carefully following the following precautions:

DO NOT short circuit.

DO NOT disassemble, deform, or apply excessive pres- sure.

DO NOT heat or incinerate.

DO NOT apply solder directly.

DO NOT mix model or new and old batteries.

DO NOT charge.

ALWAYS check that proper polarity is observed.

Remove/Replace Procedure:

Step 1. Remove the top cover and the large card-cage cover to ex- pose the A1 – A9 PCBs (refer to paragraph 8-4).

Step 2. Remove the A9 PCB from the unit.

Step 3. Locate and carefully remove the SRAM battery from the A9

PCB. See Figure 8-4.

Step 4. Replace the SRAM battery on the A9 PCB, as follows:

-Insure that the pins are clean.

-Note the polarity and carefully press the batterly into the socket.

Step 5. Reinstall the A9 PCB, the large card-cage cover, and the top cover.

Post Replacement Procedure:

Step 1. Connect power cord. Switch 373XXA power on.

Step 2. Ensure the system sweeps several times without errors.

Step 3. Refer to Chapter 3, Operational Tests. Perform the follow- ing:

8-10

373XXA MM

Page 124
Image 124
Anritsu 373XXA manual A9 PCB Sram Battery Remove and Replace