SERVICE LOG ERROR MESSAGES

APPENDIX B

Table B-3.GPIB-Related Error Messages (5 of 8)

Error Message

Description

 

 

Bad filename

The supplied filename was bad. The filename can have 8 characters maximum. No

 

extensions. The filename must start with and alpha type character (A thru Z). After

 

that the allowable characters are alpha, numeric (0 thru 9) and underscore (_).

 

 

Conflict with rotary knob

You should not be using the rotary knob and the GPIB at the same time.

 

 

Too many data points for exter-

A 6700B series external source can handle 501 data points.

nal source

A 68000 series external source can handle 999 data points.

 

 

Recalled setup corrupted

These are problems with the recalled setup.

Hardware mismatch in recalled

 

setup

 

Software mismatch in recalled

 

setup

 

 

 

Too many data points for Dis-

The maximum number of data points in discrete fill is 1601.

crete Fill

 

 

 

Not enough data points for Dis-

The minimum number or data points in discrete fill is 2.

crete Fill

 

 

 

Discrete Fill end frequency out of

The number of points for discrete fill puts the end frequency out of range.

range

 

 

 

Step is too large

When setting up a time domain harmonic sweep, cannot get 2 data points because

 

the start frequency is too high for the approximate stop frequency. In a group delay

 

display, the delay aperture percent of sweep is less than one step size.

 

 

Range too small

An attempt was made to set a distance or time span value too small. This can also be

 

done via inappropriate values for start and stop.

 

 

Start or stop out of range

An attempt was made to set a distance or time start or stop value out of range. This

 

can also be done via inappropriate values for center and span.

 

 

No bands defined

An attempt was made to turn on multiple source mode with no band definitions.

 

 

Out of frequencies for new band

The current set of multiple source mode bands use up all the frequency range of the

definition

373XXA. Therefore, no more bands can be defined.

Source out of range by equation

 

External source out of range by

 

equation

 

 

 

File is read only

An attempt was made to write to a write protected file.

 

 

File not found

An attempt was made to access a non-existent file.

 

 

Floppy drive not ready

An attempt was made to access the floppy drive with no floppy disk installed.

 

 

Floppy disk full

An attempt was made to write to a floppy disk or the hard disk when no space was left

Hard disk full

on the disk.

 

 

B-12

373XXA MM

Page 168
Image 168
Anritsu 373XXA manual Table B-3.GPIB-Related Error Messages 5

373XXA specifications

The Anritsu 373XXA series is a notable line of microwave and RF signal analyzers designed for comprehensive analysis and testing of various wireless communication technologies. These analyzers are specifically crafted to meet the demands of modern communication systems, making them ideal for engineers and technicians working in wireless research, development, and manufacturing.

One of the main features of the Anritsu 373XXA series is its wide frequency range, which typically spans from as low as 5 kHz to several GHz, accommodating a vast array of signal types and applications. This extensive frequency coverage allows users to test and validate all manner of systems, from LTE and 5G networks to satellite communications, yielding accurate and reliable measurement results.

The device employs advanced measurement technologies that provide in-depth analysis capabilities. Among these technologies is the ability to perform both time and frequency domain measurements, allowing users to visualize and diagnose signal integrity issues effectively. In addition, the 373XXA series incorporates vector signal analysis, which is crucial in evaluating complex modulated signals, commonly found in today’s sophisticated wireless communication protocols.

Another distinguishing characteristic of the Anritsu 373XXA is its user-friendly interface. It features a large, high-resolution display that enables easy navigation through the analyzer's functions and measurements. The intuitive design is further enhanced by customizable settings and measurement configurations that streamline the workflow, making it accessible for both novice and experienced users.

The 373XXA series also supports various digital and analog signal demodulation techniques, ensuring compatibility with an array of communication standards. Furthermore, users can benefit from its built-in signal generation capabilities, which allows for end-to-end testing of devices and systems under realistic operational conditions.

Portability is another key feature of the Anritsu 373XXA series. The compact design and lightweight structure facilitate easy transport, which is particularly advantageous for field applications where mobility is critical.

Overall, the Anritsu 373XXA series stands out as a versatile and powerful tool for professionals in the RF and microwave field, thanks to its extensive frequency range, advanced measurement technologies, user-friendly interface, and portability. Whether developing new wireless standards or troubleshooting existing systems, the 373XXA delivers precision and reliability essential for today’s dynamic communication landscape.