SOURCE LOCK THRESHOLD | ADJUSTMENTS |
Equipment Required
None
NOTE
Allow the 373XXA to
Procedure
Step 1. If recalibrating a 37397A , install broadband terminations to both ports or install a throughline between ports.
Step 2. Press the Option Menu key (Enhancement key group) to dis- play the OPTIONS menu.
Step 3. Select DIAGNOSTICS from menu.Then select in se- quence: H/W CALIBRATIONS then SOURCE THRESH-
OLD CALIBRATION.
Step 4. Follow the directions displayed on the 373XXA CRT screen, until calibration is completed.
Post Calibration After the calibration process is completed, perform the following ac- Actions tions, as appropriate:
If calibration passes:
Save the calibration data to (hard) disk, as follows:
Step 1. Press the Option Menu key (Enhancement key group) to dis- play the OPTIONS menu.
Step 2. Select DIAGNOSTICS from menu.Then select in se- quence: H/W CALIBRATIONS DISK OPERATIONS SAVE TO HARD DISKHW_CAL.SLT.
NOTE
Also save the calibration data to floppy disk for archival purposes as a backup in case of a hard disk failure. Label the disk with cali- bration type, date, and system model and se- rial numbers. Set the copy protect tab on the disk. Properly safeguard the disk so that it will be available for later use, as needed.
If calibration fails:
Repeat the calibration. If it still fails, then proceed to Chapter 5, Troubleshooting.
373XXA MM |