APPENDIX D

PERFORMANCE SPECIFICATIONS

Appendix D

Performance Specifications

 

SYSTEM PERFORMANCE

Frequency Range:

 

RF Models:

 

 

37317A

22.5 MHz to 8.6 GHz

Microwave Models

 

37325A

40

MHz to 13.5 GHz

37347A

40

MHz to 20 GHz

37369A

40

MHz to 40 GHz

37397A

40

MHz to 65 GHz

Dynamic Range:

The following table gives dynamic range in two manners. “Receiver Dynamic Range” is defined as the ratio of the maximum signal level at Port 2 for 0.1 dB compression to the noise floor at Port 2.

 

 

Max

Noise

Receiver

Port 1

System

Model

Freq

Signal Into

Dy-

(GHz)

Port 2

Floor

Dynamic

Power

namic

 

(dBm)

Range

(dBm)

 

 

(dBm)

Range

 

 

 

 

 

 

 

 

 

 

 

 

 

0.0225

+30

–95

125

0

95

37317A

2

+30

–98

130

0

98

 

8.6

+30

–98

130

0

98

 

0.04

+30

–65

100

+5

70

37325A

2

+30

–93

128

+5

98

 

13.5

+30

–93

128

+5

98

 

0.04

+30

–65

100

+5

70

37347A

2

+30

–93

128

+5

98

 

20

+30

–91

126

+5

96

 

0.04

+30

–65

100

+5

58

37369A

2

+30

–93

128

+5

86

20

+30

–90

125

0

83

 

 

40

+30

–83

123

–7

76

37397A

0.04

+30

–77

107

+10

70

 

2

+30

–105

135

+10

98

 

20

+30

–97

127

–2

90

 

40

+30

–95

125

–7

88

 

50

+30

–87

117

–2

80

 

65

+30

–77

107

–2

70

“System Dynamic Range” is defined as the ratio of the power incident on Port 2 in a through line connection to the noise floor at Port 2 (forward measurements only). In preparing the table, 10 Hz IF bandwidth and 512 aver- ages were used in calibration and measurement.

High Level Noise (typical)

<0.04 dB and <0.5￿ peak-to-peak variation in a 1 kHz IF bandwidth up to 20 GHz. <0.08 dB and <1.0￿ peak-to- peak variation up to 40 GHz. <0.25 dB and <2.5￿ peak-to- peak variation up to 65 GHz.

Measurement Throughput:

Measurement times are based on a single 40 MHz to 20 GHz sweep with 10 kHz IF bandwidth (no averages) after a full 12-term calibration. Sweep times include retrace and bandswitch times.

Measurement Time (ms) vs. Data Points (typical):

Calibration

 

 

Data Points

 

 

 

 

 

 

 

 

 

Type

3

51

 

101

 

401

1601

 

 

 

 

 

 

 

 

 

 

 

1 Port (3 Term)

60

250

 

330

 

960

3300

Full 2 Port

60

270

 

400

 

1000

3600

Measurement Time vs. Sweep Mode for 101 Data Points (typical):

Sweep Mode

Time (ms)

 

 

Linear

350

List

350

CW

230

Measurement Time vs. IF Bandwidth for 101 Data Points (typical):

IF Bandwidth

Time (ms)

 

 

10 kHz

350

1 kHz

530

100 Hz

1900

10 Hz

14000

373XXA MM

D-3

Page 182
Image 182
Anritsu 373XXA manual Appendix D

373XXA specifications

The Anritsu 373XXA series is a notable line of microwave and RF signal analyzers designed for comprehensive analysis and testing of various wireless communication technologies. These analyzers are specifically crafted to meet the demands of modern communication systems, making them ideal for engineers and technicians working in wireless research, development, and manufacturing.

One of the main features of the Anritsu 373XXA series is its wide frequency range, which typically spans from as low as 5 kHz to several GHz, accommodating a vast array of signal types and applications. This extensive frequency coverage allows users to test and validate all manner of systems, from LTE and 5G networks to satellite communications, yielding accurate and reliable measurement results.

The device employs advanced measurement technologies that provide in-depth analysis capabilities. Among these technologies is the ability to perform both time and frequency domain measurements, allowing users to visualize and diagnose signal integrity issues effectively. In addition, the 373XXA series incorporates vector signal analysis, which is crucial in evaluating complex modulated signals, commonly found in today’s sophisticated wireless communication protocols.

Another distinguishing characteristic of the Anritsu 373XXA is its user-friendly interface. It features a large, high-resolution display that enables easy navigation through the analyzer's functions and measurements. The intuitive design is further enhanced by customizable settings and measurement configurations that streamline the workflow, making it accessible for both novice and experienced users.

The 373XXA series also supports various digital and analog signal demodulation techniques, ensuring compatibility with an array of communication standards. Furthermore, users can benefit from its built-in signal generation capabilities, which allows for end-to-end testing of devices and systems under realistic operational conditions.

Portability is another key feature of the Anritsu 373XXA series. The compact design and lightweight structure facilitate easy transport, which is particularly advantageous for field applications where mobility is critical.

Overall, the Anritsu 373XXA series stands out as a versatile and powerful tool for professionals in the RF and microwave field, thanks to its extensive frequency range, advanced measurement technologies, user-friendly interface, and portability. Whether developing new wireless standards or troubleshooting existing systems, the 373XXA delivers precision and reliability essential for today’s dynamic communication landscape.