FREQUENCY CALIBRATION

ADJUSTMENTS

￿YIG Oscillator

￿Switched Filter

￿BBRAM chip on the A9 Processor PCB is replaced and the Source Calibration Data was not previously saved on disk (thus data could not be recalled from disk).

Calibration Perform the following steps:

Procedure

Equipment Required

Refer to Table 1-1, located on page 1-9, for further information about the following equipment:

EIP Model 578B Frequency Counter, or equivalent.

RF/Microwave Cable

GPIB cable, WILTRON 2100-2, or equivalent

NOTE

Allow the 373XXA and Frequency Counter to warm-up at least 30 minutes prior to performing calibration.

Procedure

Step 1. Press the Option Menu key (Enhancement key group) to dis- play the OPTIONS menu.

Step 2. Select DIAGNOSTICS from menu; then select in se- quence: H/W CALIBRATION, and then SOURCE FREQUENCY CALIBRATION (below).

OPTIONS

TRIGGERS

REAR PANEL

OUTPUT

DIAGNOSTICS

MULTIPLE SOURCE CONTROL

PRESS <ENTER> TO

SELECT

DIAGNOSTICS

START SELF TEST

READ SERVICE LOG

INSTALLED OPTIONS

PERIPHERAL TESTS

TROUBLESHOOTING (FOR SERVICE ONLY)

H/W CALIBRATIONS (FOR SERVICE ONLY)

PRESS <ENTER> TO

SELECT

HARDWARE CALIBRATIONS (FOR SERVICE ONLY)

LO1

CALIBRATION

LO2

CALIBRATION

SOURCE FREQUENCY CALIBRATION

SOURCE ALC

CALIBRATION

DISK OPERATIONS

PREVIOUS MENU

PRESS <ENTER> TO

SELECT

HARDWARE CALIBRATIONS (FOR SERVICE ONLY)

START ALC

FREQ CALIBRATION

PREVIOUS MENU

PRESS <ENTER> TO

SELECT

6-6

373XXA MM