Anritsu 373XXA manual Recommended Test Equipment General Information, Anritsu B39553

Models: 373XXA

1 202
Download 202 pages 15.37 Kb
Page 22
Image 22

RECOMMENDED TEST EQUIPMENT

GENERAL INFORMATION

Table 1-1.Recommended Test Equipment (1 of 2)

 

INSTRUMENT

 

CRITICAL

RECOMMENDED

USE*

 

 

SPECIFICATION

MANUFACTURER/MODEL

*

 

 

 

 

 

 

 

 

 

Computer/Controller

 

PC with Windows 95 or later and National In-

Any

O, P

 

 

 

 

struments GPIB hardware and software.

 

 

 

 

 

 

 

Test Software

 

 

ANRITSU 2300-178

O

 

 

 

 

 

Floppy Disk

 

Formatted, IBM PC format

DS/HD 1.44 Mbyte

A

 

 

 

 

 

GPIB Cable

 

 

ANRITSU 2100-2, or equivalent

O, P,

 

 

 

 

 

 

A

 

 

 

 

 

RF/Microwave Cable

 

DC – 40 GHz, K type connectors (male

Any

P, A

 

 

 

 

or female), Length, 3 ft.

 

 

 

 

 

 

 

BNC Cable

 

Length, 4 ft., 2 each

Any

O

 

 

 

 

 

Printer Port Test

 

 

ANRITSU B39553

O

Fixture

 

 

 

 

 

 

 

 

 

 

 

Thru Line

 

 

For models 37317A and below

ANRITSU 3670A50-2, 3671A50-2

O, P

 

 

 

 

For models 37325A and above:

ANRITSU 3670K50-2, 3671K50-2

 

 

 

 

 

 

Calibration Kit

 

For models 37317A and below, with Opt 7A:

ANRITSU 3651-1*

O, P

 

 

 

 

For models 37317A and below, with Opt 7N or

ANRITSU 3653

 

 

 

 

 

7NF:

 

 

 

 

 

 

For models 37325A and above:

ANRITSU 3652-1*

 

 

 

 

 

For models 37325A and below, with Opt 7S:

ANRITSU 3650-1*

 

 

 

 

 

For model 37397A

ANRITSU 3654B

 

 

 

 

 

 

Performance Verifica-

 

 

ANRITSU 2300-237

P

tion Software

 

 

 

 

 

 

 

 

 

Connector Verification

 

For models 37217B and below w/Opt 7A:

ANRITSU 3667

P

Kit

 

 

 

For models 37217B and below, with Opt 7N or

 

 

 

 

 

 

7NF:

ANRITSU 3663

 

 

 

 

 

For models 37225B and above:

 

 

 

 

 

 

For models 37247B and below, with Opt 7S:

ANRITSU 3668

 

 

 

 

 

 

ANRITSU 3666

 

 

 

 

 

For model 37397A

ANRITSU 3669B

 

 

 

 

 

 

Calibration Kit sliding load (Option {-1}), required for Performance Verification only.

 

∗∗

USE CODES:

 

 

 

 

A

Adjustment / Internal Hardware Calibration

 

 

 

O

Operational Testing

 

 

 

P

Performance Verification

 

 

 

T

Troubleshooting

 

 

 

 

 

 

 

 

 

 

1-10

373XXA MM

Page 22
Image 22
Anritsu 373XXA manual Recommended Test Equipment General Information, Anritsu B39553