Anritsu 373XXA manual A9 PCB Bbram Chip Remove and Replace, Remove/Replace Procedure

Models: 373XXA

1 202
Download 202 pages 15.37 Kb
Page 122
Image 122

A9 PCB BBRAM CHIP

REMOVE AND REPLACE

OPTIONS

TRIGGERS

REAR PANEL

OUTPUT

DIAGNOSTICS

MULTIPLE SOURCE CONTROL

PRESS <ENTER>

TO SELECT

HARDWARE CALIBRATIONS (FOR SERVICE ONLY)

LO1

CALIBRATION

LO2

CALIBRATION

SOURCE FREQUENCY CALIBRATION

SOURCE ALC

CALIBRATION

DISK OPERATIONS

PREVIOUS MENU

PRESS <ENTER> TO

SELECT

DIAGNOSTICS

START SELF TEST

READ SERVICE LOG

INSTALLED OPTIONS

PERIPHERAL TESTS

TROUBLESHOOTING (FOR SERVICE ONLY)

H/W CALIBRATIONS (FOR SERVICE ONLY)

PRESS <ENTER> TO

SELECT

H/W CALIBRATIONS DISK OPERATIONS

SAVE

TO HARD DISK

SAVE

TO FLOPPY DISK

RECALL

FROM HARD DISK

RECALL

FROM FLOPPY DISK

DELETE

FROM HARD DISK

DELETE

FROM FLOPPY DISK

COPY FROM HARD TO FLOPPY DISK

COPY FROM FLOPPY TO HARD DISK

PREVIOUS MENU

PRESS <ENTER> TO

SELECT

Preliminary:

Before removing and replacing the BBRAM Chip it is necessary to save the following calibration data to the hard disk: signal source Fre- quency, ALC, Source Lock Threshold, LO1 and LO2. Proceed as fol- lows:

Step 1. Press the Option Menu key (Enhancement key group) to display the OPTIONS menu (left).

Step 2. Select DIAGNOSTICS from the menu. Then select in se- quence: H/W CALIBRATIONS DISK OPERATIONS SAVE TO HARD DISK HW_CAL.ALL.

NOTE

Also save the calibration data to floppy disk. Label the disk with calibration type (ALL), date, and system model and serial numbers. Set the copy protect tab on the disk. Properly safeguard the disk so that it will be available for later use, as needed.

Step 3. Switch the 373XXA power off. Remove the power cord.

Remove/Replace Procedure:

Step 1. Remove the top cover and the large card-cage cover to ex- pose the A1 – A9 PCBs (refer to paragraph 8-4).

Step 2. Remove the A9 PCB from the unit.

Step 3. Locate and remove the BBRAM chip from the A9 PCB. See Figure 8-3.

Step 4. Replace the BBRAM chip on the A9 PCB.

Step 5. Reinstall the A9 PCB, the large card-cage cover, and the top cover.

Step 6. Reconnect power cord. Switch 373XXA power on.

Post Replacement Procedure:

Recall the calibration data saved in the Preliminary step from the floppy disk as follows:

Step 1. Insert the floppy disk with the previously saved Source Frequency and ALC Calibrations Data in the 373XXA floppy disk drive.

Step 2. Select DIAGNOSTICS from the menu (left). Then select in sequence: H/W CALIBRATIONS DISK OPERATIONS RECALL CAL FROM FLOPPY DISKHW_CAL.ALL.

8-8

373XXA MM

Page 122
Image 122
Anritsu 373XXA manual A9 PCB Bbram Chip Remove and Replace, Remove/Replace Procedure, Post Replacement Procedure