CAUTION:
If you continue to get control solution test results that fall outside of the range printed on the vial, the system may not be working properly. Do not use the system to test your blood if you continue to get test results that fall outside of the range. If you are unable to resolve the problem, call LifeScan Customer Care FreePhone
at UK: 0800 121200,
Republic of Ireland: 1800 535676.
Comparing control solution results
If test results fall outside the range printed on the test strip vial, repeat the test. Out-
•Error in performing the test.
•Failure to shake the control solution vial vigorously.
•Expired or contaminated control solu- tion.
•Control solution that is too warm or too cool.
•Failure to discard the first drop of con- trol solution.
•Improper coding of the meter.
•Test strip deterioration.
•Meter malfunction.