MESSAGE

WHAT IT MEANS

ACTION

 

 

 

A control solution test result

No action required.

 

 

 

 

 

 

 

 

 

stored in the memory.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

There is no test result stored in No action required. this place in the memory.

Either the ambient temperature or the meter’s temperature was too low to perform a test.

Repeat the test in a warmer place 15-35°C (59-95°F). Allow 20 minutes for the instru- ment to equilibrate to the new ambient temperature before retesting.

Either the ambient temperature or the meter’s temperature was too high to perform a test.

Repeat the test in a cooler place 15-35°C (59-95°F). Allow 20 minutes for the instru- ment to equilibrate to the new ambient temperature before retesting.