STEP 3

Apply control solution to the test strip.

To ensure an accurate control solution test, shake the vial, discard the first drop and wipe the dispenser tip before you begin.

NARROW CHANNEL

TOP EDGE

Apply Drop Here

 

Touch and hold a drop of control solution where the narrow channel meets the TOP EDGE of the test strip.

CONFIRMATION WINDOW

The control solution will be drawn into the narrow channel. When the confirmation window is full, the meter will count down from 5 to 1.

Example

The control solution test result will appear on the display. Ë will appear above your result.

17