If the meter does not display a message after inserting a test strip:

POSSIBLE CAUSE

WHAT TO DO

No power

Replace the battery.

 

 

No battery or battery

Check that the battery is

incorrectly installed

correctly installed with the positive

 

“+” side up.

 

 

Test strip inserted incor-

Insert the test strip correctly

rectly or incompletely

with the contact bars end first and

 

facing up. Push it in until it will go no

 

further.

 

 

Defective meter or test strip

Call LifeScan Customer Service

 

at 1 800 227-8862.
If the test does not start after applying the sample:

POSSIBLE CAUSE

WHAT TO DO

Insufficient blood sample

Repeat the test with a new test strip

 

and a larger sample, making sure

 

blood is placed to the narrow channel

 

in the top edge of the test strip.

 

 

Defective test strip

Repeat the test with a new

 

test strip.

Sample applied after automatic shutoff (two minutes after last user action)

Repeat the test with a new test strip; apply sample only when

the R symbol appears on the display.

Defective meter

Call LifeScan Customer Service

 

 

at 1 800 227-8862.