MESSAGE

WHAT IT MEANS

ACTION

 

 

Flashing time means the

You must reset the time, time

 

 

 

 

meter has lost power and

format and date. See Meter

 

 

entered the setting mode.

Set-up, pages 40–45. Confirm

 

 

 

that your meter is set in mg/dL.

 

 

 

If not, call LifeScan

 

 

 

Customer Service at

 

 

 

1 800 227-8862.

 

 

 

 

Note to users of OneTouch®

 

 

 

Diabetes Management

 

 

 

Software: Do Not transfer data

 

 

 

from a meter that has experi-

 

 

 

enced a loss of power. Call

 

 

 

LifeScan Customer Service

 

 

 

at 1 800 227-8862 for more

 

 

 

information.

 

 

 

 

 

 

 

The meter has detected that

Repeat the test after the meter

 

 

 

 

the temperature is below the

and test strips have reached a

 

 

system operating range. Do

temperature within the oper-

 

 

Not perform a test until the

ating range.

 

 

meter and test strips reach a

 

 

 

 

temperature within the

 

 

 

 

operating range of

 

 

 

 

43–111°F (6–44°C).

 

 

 

 

 

 

 

 

 

The meter has detected that

 

 

the temperature is above

 

 

the system operating range.

 

 

Do Not perform a test until

 

 

the meter and test strips

 

 

reach a temperature within

 

 

the operating range of

58

 

43–111°F (6–44°C).

 

 

Repeat the test after the meter and test strips have reached a temperature within the operating range.