MESSAGE

WHAT IT MEANS

 

Error message that indicates

 

 

one of the following condi-

 

tions may be present:

 

1. You may have high glu-

 

cose and have tested in an

 

environment near the low

 

end of the system’s oper-

 

ating temperature range

 

(43–111°F / 6–44°C).

 

2. There may be a problem

 

with the test strip. For

 

example, it may have been

 

damaged or moved during

 

testing.

 

3. The sample was

 

improperly applied.

56

ACTION

1.If you tested in a cool environment, repeat the test in a warmer environ- ment with a new test strip. If the error mes- sage appears again, call LifeScan

Customer Service at

1800 227-8862.

2.If you were testing in a normal or warm environ- ment, repeat the test with a new test strip. (See pages 8–9 for test strip information.) If the error message appears again, call LifeScan Customer

Service at

1800 227-8862.

3.If you applied the blood incorrectly, review pages 29–30 on blood application and repeat the test with a new test strip. If the error message appears again, call LifeScan

Customer Service at

1800 227-8862.