5.4Display and acknowledging error messages

Operation

Radar Tank Gauge

 

 

5.4Display and acknowledging error messages

Type of error

Errors that occur during commissioning or measuring are displayed immediately on the local display. If two or more system or process errors occur, the error with the highest priority is the one shown on the display.

The measuring system distinguishes between two types of error:

A (Alarm):

Instrument goes into a defined state (e.g. MAX 22 mA)

Indicated by a constant Type of error•A (Alarm):•W (Warning): symbol.

(For a description of the codes, see Table 9.2 on Page 79)

W (Warning):

Instrument continue measuring, error message is displayed.

Indicated by a flashing •E (Alarm / Warning):Error messagesManual background symbol.

(For a description of the codes, see Table 9.2 on Page 79)

E (Alarm / Warning):

Configurable (e.g. loss of echo, level within the safety distance)

Indicated by a constant/flashing Manual backgroundManual backgroundManual background symbol.

(For a description of the codes, see Table 9.2 on Page 79)

Manual backgroundManual background

– + E

Error messages

Error messages appear as four lines of plain text on the display. In addition, a unique error code is also output. A description of the error codes is given on page 79.

The "diagnostics (0A)" function group can display current errors as well as the last errors that occurred.

If several current errors occur, use O or S to page through the error messages.

The last occurring error can be deleted in the "diagnostics (0A)" function group with the function "clear last error" (0A2).

40

Installation and Operations Manual

Page 40
Image 40