Lightning Audio 7532 manual Operating conditions, Technical data, Environment, Process conditions

Models: 7532

1 106
Download 106 pages 36.55 Kb
Page 89
Image 89
Operating conditions

7532

Technical data

 

 

Inventory Control Ver- sions

All device types can be delivered as "Inventory Control Versions" with a reduced accuracy of ± 3mm (under reference conditions).

To these versions, the calibration certificate or custody transfer type approval is NOT attached.

The "Inventory Control Versions" can be selected by choosing the option »R« in feature »70« in the order code section »Custody transfer approvals« on page 8.

Operating conditions

Operating conditions

Engineering tips

See page 14

Environment

 

Ambient temperature

Ambient temperature for the transmitter:

 

range

• Standard: -40 °C … +80 °C (-40 °F … +176 °F)

 

 

• For calibration to regulatory standards: -25 °C ... +60 °C (-30

 

 

°F…+140 °F)

 

 

With Tu<-20 °C and Tu>+60 °C the operability of the LC-display is

 

 

reduced.

 

 

A weather protection cover should be used for outdoor operation if

 

 

the instrument is exposed to direct sunlight.

 

 

 

 

 

Storage temperature

-40 °C ... +80 °C (-40 °F … +176°F)

 

 

 

 

 

Climate class

DIN EN 60068-2-38 (test Z/AD)

 

 

 

 

 

Degree of protection

• housing: IP 65, NEMA 4X (open housing: IP20, NEMA 1)

 

 

• antenna: IP 68 (NEMA 6P)

 

 

 

 

 

Vibration resistance

DIN EN 60068-2-64 / IEC 68-2-64: 20…2000 Hz, 5 (m/s²)²/Hz

 

 

 

 

 

Cleaning of the antenna

See »Technical Information« TEC028

 

 

 

 

 

Electromagnetic com-

• emissions according to EN 61326; equipment class B

 

patibility

• compatibility according to EN 61326; appendix A (industrial area,

 

 

10 V/m) and Namur recommendation EMC (NE 21).

 

 

 

 

 

Process conditions

 

 

 

 

 

 

 

Process temperature

See »Technical Information« TEC028

 

range

 

 

 

 

 

 

 

Process temperature

See »Technical Information« TEC028

 

limits

 

 

 

 

 

 

 

Process pressure limits

See »Technical Information« TEC028

 

 

 

 

 

Dielectric constant

• in a stilling well: εr 1.4

 

 

• in free space: εr 1.9

 

 

 

 

 

Wetted parts

• Planar, gas-tight

 

 

Wetted parts: 1.4435 / HNBR (Hydrated Nitrit Butadien Rubber,

 

 

resistent to NH3) or FKM / PTFE glas fiber laminat

 

 

Note!

 

 

The planar antenna is not resistant to hot vapor!

 

 

• Horn adapter for increase of diameter

 

 

Wetted parts: 1.4435

 

 

 

 

 

 

 

 

89

Page 89
Image 89