9.2System error messages
Code | Description | Possible cause | Remedy |
|
|
|
|
A102 | checksum error | device has been powered off | reset; |
| general reset & new | before data could be stored; | avoid emc problem; if |
| calibr.required | emc problem; E²PROM defect | alarm prevails after reset, |
|
|
| exchange electronics |
|
|
|
|
W103 | initialising - please | E²PROM storage not yet | wait some seconds; if |
| wait | finished | warning prevails, exchange |
|
|
| electronics |
|
|
|
|
A106 | downloading please | processing data download | wait until warning |
| wait |
| disappears |
|
|
|
|
A110 | checksum error | device has been powered off | reset; |
| general reset & new | before data could be stored; | avoid emc problem; if |
| calibr.required | emc problem; E²PROM defect | alarm prevails after reset, |
|
|
| exchange electronics |
|
|
|
|
A111 | electronics defect | RAM defective | reset; |
|
|
| if alarm prevails after |
|
|
| reset, exchange electronics |
|
|
|
|
A113 | electronics defect | ROM defective | reset; |
|
|
| if alarm prevails after |
|
|
| reset, exchange electronics |
|
|
|
|
A114 | electronics defect | E2PROM defective | reset; |
|
|
| if alarm prevails after |
|
|
| reset, exchange electronics |
|
|
|
|
A115 | electronics defect | general hardware problem | reset; |
|
|
| if alarm prevails after |
|
|
| reset, exchange electronics |
|
|
|
|
A116 | download error | checksum of stored data not | restart download of data |
| repeat download | correct |
|
|
|
|
|
A121 | electronics defect | no factory calibration existant; | contact service |
|
| E²PROM defective |
|
|
|
|
|
W153 | initialising - please | initialisation of electronics | wait some seconds; if |
| wait |
| warning prevails, power off |
|
|
| device and power on again |
|
|
|
|
A155 | electronics defect | hardware problem | reset; |
|
|
| if alarm prevails after |
|
|
| reset, exchange electronics |
|
|
|
|
A160 | checksum error | device has been powered off | reset; |
| general reset & new | before data could be stored; | avoid emc problem; if |
| calibr.required | emc problem; E²PROM defect | alarm prevails after reset, |
|
|
| exchange electronics |
|
|
|
|
A164 | electronics defect | hardware problem | reset; |
|
|
| if alarm prevails after |
|
|
| reset, exchange electronics |
|
|
|
|
A171 | electronics defect | hardware problem | reset; |
|
|
| if alarm prevails after |
|
|
| reset, exchange electronics |
|
|
|
|
A231 | sensor 1 defect | HF module or electronics | exchange HF module or |
| check connection | defective | electronics |
|
|
|
|
Table 3: | System error messages |
|
|