6.4Basic Setup with the VU 331

7532

Commissioning

 

 

6.4Basic Setup with the VU 331

Function "measured value" (000)

Function "measured value" (000)6.4.1Function group "basic setup" (00)

– + E

This function displays the current measured value in the selected unit

(see "customer unit" (042) function). The number of digits after decimal point can be selected in the "no.of decimals" (095) function. The length of the bargraph corresponds to the percentile value of the present measured value with regard to the span.

6.4.1Function group "basic setup" (00)

Function "tank shape" (002)Selection:

– + E

Function "tank shape" (002)

•stilling wellManual backgroundManual background

– + E

This function is used to select the tank shape.

Note!

For the application of the 7532 RTG the selection "stilling well" has to be chosen in the function "tank shape" (002).

Selection:

dome ceiling

horizontal cyl

bypass

stilling well

flat ceiling (Typical ceiling of storage tanks: a slight slope of only a few degrees can be neglected.)

sphere

L00-FMR2xxxx-14-00-00-en-007