| Functional Details |
Setpoint configuration
You program all setpoints as part of the
There are several possible conditions (criteria) and effectively three update modes, as explained in the following configuration summary.
Set high limit
You can set the
Set low limit
You can set the
Set criteria
Inside window: Signal is below
Outside window: Signal is above
Greater than value: Signal is above
Less than value: Signal is below
Equal to value: Signal is equal to
The equal to mode is intended for use when the counter or digital input channels are the source channel.
You should only use the equal
Hysteresis mode: Outside the window, high forces output 2 until an outside the window low condition exists, then output 1 is forced. Output 1 continues until an outside the window high condition exists. The cycle repeats as long as the acquisition is running in hysteresis mode.
Set output channel
None
Update FIRSTPORTC
Update DAC
Update timerx
Update modes
Update on True only
Update on True and False
Set values for output
*By default, FIRSTPORTC comes up as a digital input. You may want to initialize FIRSTPORTC to a known state before running the input scan to detect the setpoints.
When using setpoints with triggers other than immediate, hardware analog, or TLL, the setpoint criteria evaluation begins immediately upon arming the acquisition.
Using the setpoint status register
You can use the setpoint status register to check the current state of the 16 possible setpoints. In the register, Setpoint 0 is the
37