Chapter 7 Counters
© National Instruments Corporation 7-25 NI 6124/6154 User Manual
The waveform thus produced at the counter’s output can be used to provide
timing for undersampling applications where a digitizing system can
sample repetitive waveforms that are higher in frequency than the Nyquist
frequency of the system. Figure 7-29 shows an example of pulse generation
for ETS; the delay from the trigger to the pulse increases after each
subsequent Gate active edge.
Figure 7-29. Pulse Generation for ETS
For information about connecting counter signals, refer to the Default
Counter/Timer Pinouts section.
Counter Timing SignalsSS eries devices feature the following counter timing signals:
•Counter n Source Signal
•Counter n Gate Signal
•Counter n Aux Signal
•Counter n A Signal
•Counter n B Signal
•Counter n Z Signal
•Counter n Up_Down Signal
•Counter n HW Arm Signal
•Counter n Internal Output Signal
•Counter n TC Signal
•Frequency Output Signal
In this section, n refers to either Counter 0 or 1. For example, Counter n
Source refers to two signals—Counter 0 Source (the source input to
Counter 0) and Counter 1 Source (the source input to Counter 1).
OUT
D1 D2 = D1 + ΔD D3 = D1 + 2ΔD
GATE