Chapter 6 Calibrating Images
IMAQ Vision for LabWindows/CVI User Manual 6-2 ni.com
Refer to Chapter 5, Performing Machine Vision Tasks, for more
information about applying calibration information before making
measurements.
Defining a Calibration Template
You can define a calibration template by supplying an image of a grid or
providing a list of pixel coordinates and their corresponding real-world
coordinates. This section discusses the grid method in detail.
A calibration template is a user-defined grid of circular dots. As shown in
Figure 6-1, the grid has constant spacings in the x and y directions. You can
use any grid, but follow these guidelines for best results:
The displacement in the x and y directions should be equal (dx = dy).
The dots should cover the entire desired working area.
The radius of the dots in the acquired image should be 6–10pixels.
The center-to-center distance between dots in the acquired image
should range from 18 to 32 pixels, as shown in Figure 6-1.
The minimum distance between the edges of the dots in the acquired
image should be 6 pixels, as shown in Figure6-1.
Figure 6-1. Defining a Calibration Grid
Note Click Start»Programs»NationalInstruments»Vision»Documentatio n»
Calibration Grid to use the calibration grid installed with IMAQ Vision. The dots have
radii of 2 mm and center-to-center distances of 1 cm. Depending on your printer, these
measurements may change by a fraction of a millimeter. You can purchase highly accurate
calibration grids from optics suppliers, such as Edmund Industrial Optics.
1 Center-to-Center Distance 2 Center of Grid Dots 3 Distance Between Dot Edges
dy
dx
1
3
2