Appendix B Troubleshooting
NI 6232/6233 User Manual B-2 ni.com
reference the signal to the same ground level as the device reference. There
are various methods of achieving this reference while maintaining a high
common-mode rejection ratio (CMRR). These methods are outlined in the
Connecting Analog Voltage Input Signals section of Chapter 4, Analog
Input.
AI GND is an AI common signal that routes directly to the ground
connection point on the devices. You can use this signal if you need a
general analog ground connection point to the device. Refer to the
Differential Connection Considerations section of Chapter 4, Analog
Input, for more information.
How can I use the AI Sample Clock and AI Convert Clock signals on
an M Series device to sample the AI channel(s)?
M Series devices use ai/SampleClock and ai/ConvertClock to perform
interval sampling. As Figure B-1 shows, ai/SampleClock controls the
sample period, which is determined by the following equation:
1/sample period = sample rate
Figure B-1. ai/SampleClock and ai/ConvertClock
ai/ConvertClock controls the convert period, which is determined by the
following equation:
1/convert period = convert rate
This method allows multiple channels to be sampled relatively quickly in
relationship to the overall sample rate, providing a nearly simultaneous
effect with a fixed delay between channels.
Channel 0
Channel 1
Convert Period
Sample Period