Chapter 2 DAQ System Overview
NI 6232/6233 User Manual 2-4 ni.com

Cables and Accessories

NI offers a variety of products to use with NI 6232/6233 devices, including
cables, connector blocks, and other accessories, as follows:
Cables and cable assemblies
– Shielded
Unshielded ribbon
Screw terminal connector blocks, shielded and unshielded
RTSI bus cables
For more specific information about these products, refer to ni.com.
Refer to the Custom Cabling section of this chapter, the Field Wiring
Considerations section of Chapter 4, Analog Input, and Appendix A,
Device-Specific Information, for information on how to select accessories
for your M Series device.

Custom Cabling

NI offers cables and accessories for many applications. However, if you
want to develop your own cable, the following kits can assist you:
TB-37F-37SC—37-pin solder cup terminals, shell with strain relief
TB-37F-37CP—37-pin crimp & poke terminals, shell with strain
relief
Also, adhere to the following guidelines for best results:
For AI signals, use shielded, twisted-pair wires for each AI pair of
differential inputs. Connect the shield for each signal pair to the ground
reference at the source.
Route the analog lines separately from the digital lines.
When using a cable shield, use separate shields for the analog and
digital sections of the cable. Failure to do so results in noise coupling
into the analog signals from transient digital signals.
For more information on the connectors used for DAQ devices, refer to the
KnowledgeBase document, Specifications and Manufacturers for Board
Mating Connectors, by going to ni.com/info and entering the info code
rdspmb.