NI 9144 User Guide and Specifications 30 ni.com
Table17 contains the default scan list.
NI 9205/9206 Calibration Data
The NI 9205 uses a quadratic formula for conversion from 16-bit raw data
to calibrated data.
The NI 9205 EEPROM provides overall polynominal values a3–a0 along
with gain and offset values for each voltage range, to be applied when
converting 16-bit raw data to calibrated data.
1. Convert the 32-bit hex values to 64-bit floating point format for use in
the calibration formula.
2. Select the 32-bit gain value for a particular range.
Table 17. NI 9205/9206 Scan List Format
Index Sub Type Val ue Sub Val ue
0x2001 0 ARR:U32 33 — —
1 32 — —
20x00002A38 18 0x00103238
30x00012B38 19 0x00113338