Appendix A Specifications
© National Instruments Corporation A-7 NI PXIe-6672 User Manual
Storage Environment
Ambient temperature range....................– 40 to 71 °C (Tested in
accordance with IEC-60068-2-1
and IEC-60068-2-2. Meets
MIL-PRF-28800F Class 3
low temperature limit.)
Relative humidity range.........................5% to 95% noncondensing
(Tested in accordance with
IEC-60068-2-56.)
Shock and Vibration
Operational shock.................................. 30 g peak, half-sine, 11 ms pulse
(Tested in accordance with
IEC-60068-2-27. Meets
MIL-PRF-28800F Class 2 limits.)
Random vibration
Operating........................................ 5 to 500 Hz, 0.3 grms
Nonoperating.................................. 5 to 500 Hz, 2.4 grms
(Tested in accordance with
IEC-60068-2-64. Nonoperating
test profile exceeds the
requirements of
MIL-PRF-28800F, Class 3.)
Note Specifications are subject to change without notice.
Safety
This product is designed to meet the requirements of the following
standards of safety for electrical equipment for measurement, control, and
laboratory use:
• IEC 61010-1, EN 61010-1
• UL 61010-1, CSA 61010-1
Note For UL and other safety certifications, refer to the product label or visit ni.com/
certification, search by model number or product line, and click the appropriate link
in the Certification column.