Chapter 9 | Counters |
Counter n Gate Signal
The Counter n Gate signal can perform many different operations depending on the application including starting and stopping the counter, and saving the counter contents.
Routing a Signal to Counter n Gate
Each counter has independent input selectors for the Counter n Gate signal.
Any of the following signals can be routed to the Counter n Gate input.
•PFI <0..3>, PFI <8..11>
•ai/ReferenceTrigger
•ai/StartTrigger
•ai/SampleClock
•ai/ConvertClock
•ao/SampleClock
In addition, Counter 1 Internal Output or Counter 1 Source can be routed to Counter 0 Gate. Counter 0 Internal Output or Counter 0 Source can be routed to Counter 1 Gate.
Some of these options may not be available in some driver software.
Routing Counter n Gate to an Output Terminal
You can route Counter n Gate out to any PFI <4..7> or PFI <12..15> terminal.
Counter n Aux Signal
The Counter n Aux signal indicates the first edge in a
Routing a Signal to Counter n Aux
Each counter has independent input selectors for the Counter n Aux signal.
Any of the following signals can be routed to the Counter n Aux input.
•PFI <0..3>, PFI <8..11>
•ai/ReferenceTrigger
•ai/StartTrigger
In addition, Counter 1 Internal Output, Counter 1 Gate, Counter 1 Source, or Counter 0 Gate can be routed to Counter 0 Aux. Counter 0 Internal