6
Connecting AI Signals on the
USB-6215/6218 Devices
You can connect the
When connecting various sources to the
Note You must always connect AI GND to a local ground signal in your system using a low impedance connection. If you leave AI GND unconnected, you cannot ensure that AI <0..31> are within 10 V of AI GND, and your measurement may be unreliable.
Differential Measurements
To attain more accurate measurements and less noise, use a differential measurement configuration. A differential measurement configuration requires two inputs for each measurement.
Differential Pairs
Table
| Table | I/O Connector Signals |
| ||
|
|
|
|
| |
| Channel |
| Signal + |
| Signal – |
|
|
|
|
|
|
|
|
|
|
|
|
0 |
| AI 0 |
| AI 8 |
|
|
|
|
|
|
|
1 |
| AI 1 |
| AI 9 |
|
|
|
|
|
|
|
2 |
| AI 2 |
| AI 10 |
|
|
|
|
|
|
|