Chapter 11 Isolation and Digital Isolators
Switching voltages through the transformer cause charging and discharging of the parasitic capacitances and inductances in the switching power supplies that occur on every switch cycle, resulting in
These parasitic currents interact with parasitic and
To reduce
•Better grounding from the front connector AI GND to the signal source ground can reduce
•Reduce source impedances if possible. The parasitic currents react with these impedances.
Creating an AC Return Path
Caution Adding a capacitor will degrade the USB M Series device withstand voltage and isolation specifications. Withstand voltage must be retested by an approved testing facility after adjustments are made to the measurement system.
Isolated Systems
A fully isolated measurement system is one where the device’s isolated front end is not connected back to earth ground.
Create an AC path back to earth ground from the device’s isolated ground by connecting a high voltage capacitor between the isolated board ground and earth ground. The voltage rating of the capacitor must be larger than the voltage drop between the isolated ground and earth ground.
A
© National Instruments Corporation |