© National Instruments Corporation 21 NI WLS/ENET-9163 Carrier User Guide and Specifications
Note Pause triggers are only sensitive to the level of the source, not the edge.
Analog Input Timing SignalAI Sample Clock
A sample consists of one reading from each channel in the AI task.
ai/SampleClock signals the start of a sample of all analog input channels
in the task. ai/SampleClock can be generated from external or internal
sources.
Using A Digital Source
To use ai/SampleClock, specify a source and an edge on PFI 1. For more
information, refer to the Device Routing in MAX topic in the NI-DAQmx
Help.
Figure 11. Sample Clock Timing Options

Routing AI Sample Clock to an Output Terminal

You can route ai/SampleClock to the PFI 1 terminal.
Convert Behavior For Analog Input ModulesScanned
Scanned C Series analog input modules contain a single A/D converter and
a multiplexer to select between multiple input channels. When the C Series
Module Interface receives a Sample Clock pulse, it begins generating a
Convert Clock for each scanned module in the current task. Each Convert
Clock signals the acquisition of a single channel from that module. The
Convert Clock rate depends on the module being used, the number of
channels, and the Sample Clock rate.
Programmable
Clock
Divider
ai/SampleClock Timebase
20 MHz Timebase
PFI
ai/SampleClock
Sigma-Delta Module Internal Output