© National Instruments Corporation 21 NI WLS/ENET-9163 Carrier User Guide and Specifications
Note Pause triggers are only sensitive to the level of the source, not the edge.
Analog Input Timing SignalAI Sample ClockA sample consists of one reading from each channel in the AI task.
ai/SampleClock signals the start of a sample of all analog input channels
in the task. ai/SampleClock can be generated from external or internal
sources.
Using A Digital Source
To use ai/SampleClock, specify a source and an edge on PFI 1. For more
information, refer to the Device Routing in MAX topic in the NI-DAQmx
Help.
Figure 11. Sample Clock Timing Options
Routing AI Sample Clock to an Output Terminal
You can route ai/SampleClock to the PFI 1 terminal.
Convert Behavior For Analog Input ModulesScannedScanned C Series analog input modules contain a single A/D converter and
a multiplexer to select between multiple input channels. When the C Series
Module Interface receives a Sample Clock pulse, it begins generating a
Convert Clock for each scanned module in the current task. Each Convert
Clock signals the acquisition of a single channel from that module. The
Convert Clock rate depends on the module being used, the number of
channels, and the Sample Clock rate.
Programmable
Clock
Divider
ai/SampleClock Timebase
20 MHz Timebase
PFI
ai/SampleClock
Sigma-Delta Module Internal Output