Omega 3200ST, 3400ST Troubleshooting Guide, Symptom Possible Cause Method of Correction

Models: 3200ST 3200 3400 3400ST

1 26
Download 26 pages 61.63 Kb
Page 23
Image 23

H.Troubleshooting Guide

Symptom

Possible Cause

Method of Correction

 

 

 

Clicking noise from

Unit in error mode

Check there is sufficient

controller

 

pressure and that the flow

 

 

path is not restricted or

 

 

blocked

 

 

 

No response

Unit wired incorrectly

Check wiring is according to

 

 

Section B5

 

 

 

 

Loose connection

Check all connectors and

 

 

wiring

 

 

 

 

Damaged connector pins

Contact Omega Engineering

 

 

 

 

Blocked flow path

Check flow path for

 

 

obstructions.

 

 

 

 

Piping leak before sensor

Check all piping and

 

 

connections.

 

 

 

 

Insufficient power

Check the power supply

 

 

output and increase if

 

 

necessary

 

 

 

 

Output load resistance too low

Ensure the voltmeter or

 

 

data acquisition system or

 

 

display has an impedance of

 

 

greater than of 2.5kohm

 

 

 

 

Flow too low for the unit

Ensure that the flow being

 

 

measured is within the

 

 

capabilities of the unit

 

 

 

 

Unit damaged or faulty

Contact Omega Engineering

 

 

 

Inaccurate control

Particles in flow path

Add filtration before the

 

 

sensor.

 

 

 

 

Flow path obscured

Remove any debris or

 

 

blockage in the flow path

 

 

eg. PTFE tape.

 

 

 

 

Unit calibrated for a different

Check calibration certificate

 

gas

and apply a “K” Factor to

 

 

readings if necessary.

 

 

 

 

Gas composition is variable

Contact Omega Engineering

 

 

 

 

Fittings have been changed

Replace the factory installed

 

 

fittings

 

 

 

 

Moisture in gas

Ensure gas is clean and dry

 

 

 

 

Insufficient warm-up period

Allow the unit to warm-up

 

 

for at least 5 minutes.

 

 

 

 

Zero drift

Verify the zero and adjust

 

 

as necessary as explained in

 

 

Section C

 

 

 

 

The gain potentiometer has

Contact Omega Engineering

 

been adjusted

 

 

 

 

 

Unit needs recalibration

Contact Omega Engineering

 

 

 

 

Flow too high for the unit

Ensure that the flow being

 

 

measured is within the

 

 

capabilities of the unit

 

 

 

M-4271/0707, pg. 23 of 26

Page 23
Image 23
Omega 3200ST, 3400ST manual Troubleshooting Guide, Symptom Possible Cause Method of Correction