2.Mounting the Flow Controller.

The FMA 3200/3400 Series controllers have no particular orientation or installation requirements so may be mounted in any convenient position.

It is recommended that units be fixed to a suitable substrate using the two 4-40 mounting holes provided.

Mounting View from Bottom

(mounting hardware not included with sensor)

3.Tubing Connections

All tubing must be clean, dry and purged with clean dry air before installation of the FLO-CONTROLLER®.

If the gas to be used may contain particles then a filter (20 microns or less) should be installed upstream of (before) the unit.

When connecting the sensor to the tubing, take care not to over-tighten the fittings or leaking may occur.

Caution: Only use the fittings factory installed on the unit. If the fittings are removed the calibration of the unit may be effected and leaking may occur. If different fittings are required please contact the Omega Engineering Customer Service Department for assistance.

M-4271/0707, pg. 7 of 26

Page 7
Image 7
Omega 3200ST, 3400ST manual Mounting the Flow Controller, Tubing Connections