![* Metric mL per Revolution numbers are rounded off](/images/new-backgrounds/35301/3530185x1.webp)
Specifications 9
| Table |
| ||
|
|
|
|
|
Tubing Wall | Tubing Size | mL per | Minimum Flow | Maximum Flow |
Thickness | Inner Diameter | Revolution | Rate at 600 RPM | Rate at 600 RPM |
|
|
| (mL/Minute) | (mL/Minute) |
1.5 mm | 3.0 mm | 0.8* | 8 | 482 |
1.5 mm | 4.0 mm | 1.43* | 15 | 857 |
1.5 mm | 5.0 mm | 2.1* | 21 | 1257 |
1.5 mm | 6.0 mm | 2.7* | 27 | 1607 |
1.5 mm | 7.0 mm | 3.6* | 37 | 2187 |
1.5 mm | 8.0 mm | 3.9* | 39 | 2316 |
* Metric mL per Revolution numbers are rounded off
Normalized flow rates (mL per revolution) vary significantly, based on motor speed, tubing materials, and mechanical tolerances in pump dimensions. In addition, the variances are different for the different tubing dimensions. Variances from the normalized flow rates of Table
Tubing operating life tests are done at 600 RPM, with 20°C water, 0 PSI back pressure until the tubing breaks
Table
Tubing Size | Due to | Between Pumps | Due to Different |
Inner Diameter | Motor | (due to Mechanical | Tubing Materials |
(in.) | Speed | Tolerances) |
|
|
|
|
|
1⁄32 | ±10% | ±25% | ±20% |
1⁄16 | ±10% | ±15% | ±15% |
1⁄8 | ±10% | ±10% | ±5% |
3⁄16 | ±5% | ±10% | ±5% |
1⁄4 | ±5% | ±15% | ±5% |
5⁄16 | ±15% | ±20% | ±10% |