Guidance and manufacturer’s declaration – electromagnetic immunity – for all EQUIPMENT and

SYTEMS:

The monitor is intended for use in the electromagnetic environment specified below.The customer or the user of the monitor should assure that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment –

guidance

 

 

 

Electrostatic

6 kV contact

6 kV contact

Floors should be wood, concrete or

discharge (ESD)

8 kV air

8 kV air

ceramic tile. If floors are covered with

IEC 61000-4-2

synthetic material, the relative humidity

 

 

 

 

should be at least 30%.

 

 

 

 

 

 

 

Electrical fast

2 kV for power supply

2 kV for power supply

Mains power quality should be that

transient/burst

lines

lines

of a typical commercial or hospital

IEC 61000-4-4

1 kV for input/output

1 kV for input/output

environment.

 

 

lines

lines

 

 

 

 

 

Surge

1 kV line(s) to line(s)

1 kV line(s) to line(s)

Mains power quality should be that

IEC 61000-4-5

2 kV line(s) to earth

2 kV line(s) to earth

of a typical commercial or hospital

environment.

 

 

 

interruptions and

<5 % UT (>95 % dip in

<5 % UT (>95 % dip in

Mains power quality should be that

voltage variations

UT) for 0.5 cycle

UT) for 0.5 cycle

of a typical commercial or hospital

on power supply

40 % UT (60 % dip in

40 % UT (60 % dip in

environment. If the user of the monitor

input lines

requires continued operation during

UT) for 5 cycles

UT) for 5 cycles

IEC 61000-4-11

power mains interruptions, it is

 

 

70 % UT (30 % dip in

70 % UT (30 % dip in

recommended that the monitor be

 

 

UT) for 25 cycles

UT) for 25 cycles

powered from an uninterruptible power

 

<5 % UT (>95 % dip in

<5 % UT (>95 % dip in

supply or a battery.

 

 

 

UT) for 5 sec

UT) for 5 sec

 

 

 

 

 

Power frequency

3 A/m

3 A/m

Power frequency magnetic fields should

(50/60 Hz)

 

 

be at levels characteristic of a typical

magnetic field

 

 

location in a typical commercial or

IEC 61000-4-8

 

 

hospital environment.

 

 

 

 

 

 

 

Note

UT is the a.c. mains voltage prior to application of the test level.