Recommended separation distances between portable and mobile RF communications equipment and the EQUIPMENT or SYSTEM – for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING:

The monitor is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.The customer or the user of the monitor can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the monitor as recommended below, according to the maximum output power of the communications equipment.

 

Separation distance according to frequency of transmitter (Meter)

Rated maximum output

 

 

 

150 kHz to 80 MHz

80 MHz to 800 MHz

800 MHz to 2.5 GHz

power of transmitter (W)

d = 1.2

d = 1.2

d = 2.3

 

 

 

 

 

0.01

0.12

0.12

0.23

 

 

 

 

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

 

 

 

 

10

3.8

3.8

7.3

 

 

 

 

100

12

12

23

 

 

 

 

Note

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.