Gettin
g
Familair with the Power Supply 4 - 11
9) Disable the overcurrent protection and reset the output channel by pressin
g
:
Check that the display reads approximately 0.5A.
Also check that the CC annunciator is on, indicatin
g
that the output channel
is in the ’Constant Current’ mode a
g
ain.
10) Turn off the power supply.
11) Remove the jumper wire between the - V(olta
g
e) and + V(olta
g
e) output
terminals and reattach the terminal block cover.
RESET
OCP EN
OCP DIS