Calibrating

Calibrating Outputs

1.Select Options > Calibrate > Reset to set the generator’s calibration factors to default nominal values.

2.Press Options > Calibrate > Full Scale.

3.Set the voltmeter’s scale factor to be able to measure a nominal 1000 millivolts DC.

4.Using a VGA-to-BNC cable, connect the generator’s red output to the positive input on the voltmeter, with a 75 ohm (+/- 1%) input terminator. Connect the negative lead of the meter to any ground pin.

5.Select R_FS+ or -R_FSto set the DC output level to 1000 mV +/- 3 mV.

6.Repeat steps 4 and 5 for the green and blue outputs.

7.Select Save.

8.Select Options > Calibrate > Zero.

9.Connect the red output to the positive input of the voltmeter.

_____ Calibrate

_

 

_Full_Scale____

Zero

 

_Reset

 

 

 

 

 

 

 

 

 

 

 

 

 

 

_R_FS+_____

___-R_FS

 

_G_FS+_______

_-G_FS

 

_B_FS+_______

_-B_FS

 

_Save_________

_Back

 

 

 

 

 

 

 

 

 

 

_R_Zero+____

-R_Zero

 

_G_Zero+____

-G_Zero

 

_B_Zero+____

-B_Zero

 

_Save__________

Back

 

 

 

10.Select R_Zero+ or -R_Zeroto set the DC output level to 0 mV +/- 3 mV.

11.Repeat steps 9 and 10 for the green and blue components.

12.Select Save.

13.Select Options > NTSC_CAL.

14.Adjust the oscilloscope's timebase and sweep trigger to view the entire waveform. Adjust the G_NTSC so that its chrominance is 100 IRE or 714mV.

15.Using a vector scope, use the R_NTSC and B_NTSC options to adjust the red and green outputs using so that their signals align with the 2% boxes.

16.Press Save.

Note: The modulated RF output does not have any calibration settings.