Calibrating
Calibrating Outputs
1.Select Options > Calibrate > Reset to set the generator’s calibration factors to default nominal values.
2.Press Options > Calibrate > Full Scale.
3.Set the voltmeter’s scale factor to be able to measure a nominal 1000 millivolts DC.
4.Using a 
5.Select R_FS+ or 
6.Repeat steps 4 and 5 for the green and blue outputs.
7.Select Save.
8.Select Options > Calibrate > Zero.
9.Connect the red output to the positive input of the voltmeter.
_____ Calibrate
_
  | _Full_Scale____  | Zero  | 
  | _Reset  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | _R_FS+_____  | |
  | _G_FS+_______  | |
  | _B_FS+_______  | |
  | _Save_________  | _Back  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | _R_Zero+____  | |
  | _G_Zero+____  | |
  | _B_Zero+____  | |
  | _Save__________  | Back  | 
  | 
  | 
  | 
10.Select R_Zero+ or 
11.Repeat steps 9 and 10 for the green and blue components.
12.Select Save.
13.Select Options > NTSC_CAL.
14.Adjust the oscilloscope's timebase and sweep trigger to view the entire waveform. Adjust the G_NTSC so that its chrominance is 100 IRE or 714mV.
15.Using a vector scope, use the R_NTSC and B_NTSC options to adjust the red and green outputs using so that their signals align with the 2% boxes.
16.Press Save.
Note: The modulated RF output does not have any calibration settings.