Technical specifications

General technical specifications

Power supply

90W power supply unit

 

Input range: 100–240V; 50–60Hz; 115V, 400Hz nominal for aircraft use

 

Typical power consumption: 70W (80VA)

 

Maximum power consumption: 110W (120VA)

 

90W DC/DC converter

 

Nominal inputs: 12V, 24V

 

Typical power consumption: 70W

 

Maximum power consumption: 110W

Environmental

Operating temperature: +41ºF to +95ºF (+5ºC to +35ºC)

conditions

Note: The air flow for breathing produced by this therapy device can be higher

 

than the temperature of the room. Under extreme ambient temperature conditions

 

(104°F/40°C) the device remains safe.

 

Operating humidity:10 to 95% non-condensing

 

Operating altitude: Sea level to 8,500’ (2,591 m); air pressure range 1013 hPa to

 

738 hPa

 

Storage and transport temperature: -4ºF to +140ºF (-20ºC to +60ºC)

 

Storage and transport humidity: 10 to 95% non-condensing

Electromagnetic

Product complies with all applicable electromagnetic compatibility requirements

compatibility

(EMC) according to IEC60601-1-2, for residential, commercial, and light industry

 

environments.

 

It is recommended that mobile communication devices are kept at least one

 

meter away from the device.

 

Information regarding the electromagnetic emissions and immunity of these

 

ResMed devices can be found on www.resmed.com, on the Products page under

 

Service and Support. Click on the PDF file for your language.

Aircraft use

ResMed confirms that the S9 VPAP Tx meets the Federal Aviation Administration

 

(FAA) requirements (RTCA/DO-160, section 21, category M) for all phases of air

 

travel.

IEC 60601-1

Class II (double insulation), Type BF, Ingress Protection IP21

classification

 

Measuring and

Pressure sensor: Internally located at device outlet, analog gauge pressure type,

display devices

-5 to +45 cm H2O

 

Flow sensor: Internally located at device inlet, digital mass flow type,

 

-70 to +200 L/min

 

S9 VPAP Tx technical specifications

Mode pressure

CPAP, AutoSet mode

ranges

Pressure: 4–20 cm H2O;

 

S, ST, T and PAC mode

 

IPAP: 4–30 cm H2O; EPAP: 2–25 cm H2O;

 

iVAPS mode

 

PS: 0–27 cm H2O; EPAP: 2–25 cm H2O;

 

VAuto mode

 

PS: 0-10 cm H2O; Min EPAP: 4 cm H2O; Max IPAP: 25 cm H2O;

 

ASV and ASVAuto mode

 

PS: 0–20 cm H2O; EPAP: 4–15 cm H2O;

Maximum single

Maximum single fault steady state pressure: 30 cm H2O—if pressure exceeded

fault pressure

for > 6 sec; 40 cm H2O—if pressure exceeded for >1 sec